Abstract

A flow injection analysis (FIA) procedure for the speciation of Cr(III) and Cr(VI) using the 1, 5-diphenylcarbazide (DPC) method is presented. As Cr(III) does not interfere in the Cr(VI)-DPC reaction, both Cr(VI) and total chromium [after the on-line oxidation of Cr(III) by Ce(V)] are sequentially determined. Cr(III) is obtained by difference. Under the experimental conditions described, the calibration graphs are linear up to 2 µg ml–1 of Cr(VI) and 4 µg ml–1 of Cr(III). The detection limits found were 18 ng ml–1 for Cr(VI) and 55 ng ml–1 for Cr(III), at a signal to noise ratio of 3. The common interfering elements in the Cr(VI)-DPC reaction were investigated under dynamic FIA conditions. The FIA method was also compared with the conventional spectrophotometric procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.