Abstract

In cementless total hip arthroplasty, the rotational stability of the stem is generally confirmed in a one-time manual evaluation by the surgeon. This study was performed to evaluate the sequential intraoperative changes in rotational stability using a torque wrench. Primary total hip arthroplasty was performed on 52 consecutive hips using a single wedge stem design. Intraoperative evaluation of rotational stability was uniformly performed using a torque wrench. Evaluations were performed immediately after broach insertion and immediately before the final insertion of the stem (after placement of the acetabular cup). Immediately after the insertion of the broach, rotational stability was checked and confirmed to be fully stabilized using a torque wrench in all cases, and the stability was maintained in 17 of 52 (33%) hips immediately before the final insertion of the stem. Among the hips showing instability, 11 of 35 (31%) broaches were upsized, while the remaining 24 hips achieved stabilization through deeper insertion of the broach. In conclusion, the rotational stability achieved immediately after the insertion of the broach was not necessarily maintained during surgery, indicating that rotational stability may need to be checked at multiple time points intraoperatively.