Abstract

A sequential injection system for phosphate determination within a wide concentration range was developed for water samples with high salinity levels. The determination is based on molybdenum blue chemistry using the standard addition quantification method. The detection system included a multi-reflective flow cell coupled to a light emitting diode, enabling the minimization of the schlieren effect. The developed system gave a LOD of 0.3 μmol P L−1 and LOQ of 1.1 μmol P L−1 with a sample consumption of 125 μL. A determination rate of about 30 h−1 was obtained, and the developed method was effectively applied to interstitial, transitional and coastal waters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.