Abstract

The monitoring of the presence of plutonium isotopes is a very important issue in both nuclear energy industry and environmental protection for human health risk assessment. The developed method for the determination of 239Pu isotope with ICP-QMS detection is based on the use of sequential injection analysis (SIA) system for the pretreatment of samples to be analyzed with the use of a commercial sorbent called diglycolamide (DGA-b) for the analyte preconcentration and separation from 238U, which is the source of the most significant isobaric interference 238U1H+. Optimizing the size of the sorbent bed and the conditions of analyte sorption and elution of plutonium and uranium provides the analytical procedure for 239Pu determination, where at total analysis time of 45 min (at flow rate 10 mL min−1) the evaluated method detection limit (MDL) for 100 mL sample volume was 88 mBq L−1. This is satisfactory detectability for applications in both technological and environmental samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.