A flow-injection method has been developed for the sequential spectrophotometric determination of iron and titanium using 3,4 dihydroxybenzoic acid as chromogenic reagent. The system involves the sequential measurement of the absorbances of the complexes at 380 and 570 nm. The system is designed using a simultaneous injection of sample and reagent into separate carrier streams. The proposed method is characterized by a precision of about 2%, a sampling rate of about 50 samples per hour, and a reagent consumption of 200 μl (0.50% solution) per sample. It is relatively free of interferences and was used for the sequential determination of titanium and iron in rocks.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.