Abstract

In the area of instrument evaluation, one aspect that still requires objective assessment is the dynamics of instrument maneuver and exchange. If we could gain a better understanding of these phenomena, we could improve the design of the instruments themselves. A total of 29 laparoscopic procedures were videotaped and reviewed using time motion analysis. Instrument multifunctionality was determined using a standardized list of laparoscopic maneuvers. State transition diagrams were utilized to document the sequence of instrument exchanges. The curved dissector, atraumatic grasper, and cautery scissors were identified as the most multifunctional instruments; each was able to perform five distinct maneuvers. Instrument sequences were found to consist of a three-part dissect --> clip --> cut cycle and a two-part dissect --> suction cycle of instrument exchange. This study demonstrated that laparoscopic instruments are often used to perform a variety of maneuvers in addition to their primary function. Furthermore, there are common patterns in instrument exchange that provide a potential source of design parameters for improved surgical efficiency.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.