Abstract

A new chelating resin, 2-mercaptobenzothiazole loaded Amberlite XAD-1180 was prepared and used for separation and preconcentration of Cd(II) and Pb(II) ions prior to their determinations by flame atomic absorption spectrometry. The optimum pH for simultaneous retention of the elements and the best elution means for their simultaneous elution were 9.5 and 2molL−1 HNO3, respectively. The detection limits for Cd(II) and Pb(II) were 0.35 and 5.0μgL−1, respectively. The accuracy of the method was confırmed both by analyzing the certified reference material (RM 8704 Buffalo river sediment) and performing recovery studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.