Abstract

AbstractA new sequential continuous chromatographic refiner (SCCR‐2) for high temperature production scale G.L.C. (gas‐liquid chromatographic) separations is described. In this equipment the counter‐current movement between the gas and liquid phase is simulated by sequencing a system of inlet and outlet port functions around twelve static 2.54 cm internal diameter and 61 cm long stainless steel columns. The versatility of the equipment has been demonstrated by the separation of equivolume mixtures of the halocarbons arklone P/genklene P, methylchloroacetate/ethyl lactate and ethyl caprate/ethyl laurate at temperatures of 60°, 105° and 160°C respectively. Throughputs between 21–75 cm3h−1 have been explored with product purities in excess of 99.8% achieved under certain process conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.