Abstract

Abstract We have developed a separation method of alpha-ray and gamma-ray images using the imaging plate (IP). The IP from which the first image was read out by an image reader was annealed at 50°C for 2 h in a drying oven and the second image was read out by the image reader. It was found out that an annealing ratio, k , which is defined as a ratio of the photo-stimulated luminescence (PSL) density at the first measurement to that at the second measurement, was different for alpha rays and gamma rays. By subtracting the second image multiplied by a factor of k from the first image, the alpha-ray image was separated from the alpha and gamma-ray mixed images. This method was applied to identify the images of helium, carbon and neon particles of high energies using the heavy-ion medical accelerator, HIMAC.