A reversed phase gradient ultra performance liquid chromatography tandem mass spectrometry (UPLC/MS/MS) method has been developed for the analysis of smokeless powders. A total of 20 different components were separated by UPLC and detected by MS/MS in multiple reaction monitoring (MRM) mode. These compounds included diphenylamines, centralites, nitrotoluenes, nitroglycerin, and various phthalates. Simultaneous positive and negative electrospray ionization (ESI) was used along with negative atmospheric pressure chemical ionization (APCI) to detect all compounds in a single analysis. Analysis times were under 8min with a gradient of 10-73% organic at a flow rate of 0.500mL/min. With this method, ultraviolet and MRM limits of detection ranging from 0.08 to 2.6ng and 0.4-64ng injected were achieved. Commercially available smokeless powders were also extracted with methylene chloride and characterized using the developed UPLC/MS/MS method. The procedure permits the determination of compositional differences between different brands as well as lot-to-lot variations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.