Abstract
A number of products are available to dentists for the rapid production of intraoral radiographic films but there is little information on their relative merits. This study evaluated the performance of five "rapid" film processing products commonly used by British dentists in comparison with standard Kodak manual processing. Two Perspex contrast-detail test objects were made in order to investigate threshold contrast. Film speed, film gradient, limiting resolution and threshold contrast results are presented. Rapid processing systems possessed lower film speed when compared with Kodak Ektaspeed film and standard Kodak processing. The speed of E-speed film was found to be lower than that of D-speed film when used with Westone "Rapid X-ray" processing. Overall image quality was generally similar for all film/processing combinations evaluated, with the exception of Nix QP which gave markedly poorer image quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.