Abstract
A determination of the sensitometric parameters of screen-film systems to evaluate their qualities was performed. The quality control of the automatic film processor was carried out to ensure a high level of efficiency. Based on ISO 9236-3, the following potentials were applied on the X-ray tubes: 25 kV, 28 kV, 30 kV and 35 kV. Four different mammography films from different manufacturers with and without screens were tested for curve shape, speed and average gradient. The results indicated that film 1 exhibited better contrast, film 3 demonstrated the highest energy dependence, and film 4 presented the largest base+fog density. None of the four mammographic films tested achieved satisfactory results in all parameters analyzed. Improvements in the manufacturing process for these films must be completed to avoid losses in the image quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.