Abstract
The sensitivity of X-ray phase imaging based on Talbot interferometry is discussed. In order to evaluate the superiority of the technique to the conventional absorption-contrast method, which relies on X-ray absorption, a criterion is proposed. An experimental result of X-ray phase imaging with a Talbot interferometer is compared with the criterion. The criterion is also available for X-ray phase imaging based on Talbot–Lau interferometry. The advantage of X-ray phase imaging based on Talbot(–Lau) interferometry is more prominent when smaller structures are observed with smaller pixels.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.