Abstract

The integral quality monitor (IQM) transmission detector is a wedge-shaped large area ionization chamber that reports a position-weighted dose area product for each control point of an IMRT or VMAT plan. In this study, the accuracy of the signal prediction is verified for the Synergy Agility MLC. Tolerance criteria for VMAT plan verification with the IQM were obtained from the observed sensitivity for the detection of incorrectly delivered plans. The predicted IQM signal was compared to the measured signal recorded for a set of 30 VMAT plans for each beam quality of 6 and 10MV. The system's capability to detect incorrectly delivered plans was tested by measuring altered plans containing small, random deviations. In addition, the observed deviations were related to measurements performed with a second QA phantom. The cumulative IQM signal per arc deviated from the respective calculation on average by -0.48% (6MV) and +0.21% (10MV) with a standard deviation of 1.08% in both cases, suggesting a 2% warning and 3% action threshold as plan acceptance criteria. This choice was confirmed by the optimum threshold of 2.5% obtained via receiver operating characteristic (ROC) analysis. Reproducibility of individual control points in multiply measured plans was low (on average 7% for 1SD) and thus, segment-by-segment comparison was impractical. A suitable criterion to resolve the angular distribution of the plan was identified by binning three to five control points as a running average. While the correlation between IQM signal deviations and gamma passing rates obtained with the ArcCHECK phantom was low for clinical plans, it was apparent for erroneous plans. Binning led to even higher sensitivity to errors. The IQM was able to detect induced errors at least as reliable as the standard phantom and showed the potential to be used in pretreatment plan verification to ensure the correct plan transfer and delivery. However, there is no direct correlation between the IQM signal deviation and DVH metrics, so the IQM should be primarily used to screen for errors. Finer diagnostics should then be carried out using a different phantom.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.