Abstract

To evaluate the sensitivity of T2-weighted fast spin-echo (FSE) sequences to physiological iron depositions in normal brains at MR imaging field strengths of 1.5 and 3.0 T. T2-weighted FSE sequences acquired at 1.5 and 3.0 T clinical imaging systems (Gyroscan Intera, Philips Medical Systems, Best, The Netherlands) were compared by means of MRI in phantoms ( n=6) and healthy volunteers ( n=10). Contrast-to-noise ratios (CNRs) of tubes doped with iron oxides at different concentrations and of brain areas with physiological iron depositions (nucleus ruber, substantia nigra, globus pallidus) were calculated for either field strength. Apparent susceptibility effects of iron-containing brain structures were qualitatively analyzed by comparing the degree of visible hypointensity by a score system at either field strength. The mean CNR of iron oxide tubes and iron-containing brain areas was significantly decreased at 3.0 T. Qualitative analysis confirmed these measurements. Detection and diagnosis of brain disorders with altered iron content such as neurodegenerative parkinsonian disorders (NPD) or intracerebral hemorrhage should benefit from the increased sensitivity of T2-weighted FSE sequences to susceptibility effects at 3.0 T.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.