Abstract

Treatment planning systems that use the Monte Carlo algorithm can calculate the dose to the medium (Dm) in non-water-equivalent tissues such as bones. However, Dm cannot be verified using actual measurements; therefore, it is necessary to develop tissue-equivalent dosimeters. In this study, we developed a bone-equivalent polymer gel dosimeter (BPGD) that can measure the dose absorbed by the bone and investigated its sensitivity. The BPGDs were prepared by adding 3.0mol of calcium hydrogen phosphate dihydrate as a component of bone to an improved dose-sensitive polyacrylamide gelatin and tetrakis hydroxymethyl phosphonium chloride (iPAGAT). One day after preparation, the BPGDs were irradiated with a field size of 15 × 15 cm2 using a 10 MV X-ray beam to evaluate the dose sensitivity, dose-rate dependence, and dose-integration dependence. One day after dose exposure, the BPGDs were scanned using a 0.4T MRI APERTO Eterna (Hitachi, Tokyo, Japan) to obtain R2 values. The difference between the R2 values of 6Gy and 0Gy was up to 5s-1, and the R2 curve plateaued in the high-dose region. Moreover, the BPGD did not depend on the integration of the dose and dose rates. Therefore, the BPGDs that we developed can determine the radiation dose to bones.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.