Abstract

We present a study of the uncertainties on the measures of the surface resistance, obtainable with different measurement methods based on the use of dielectric-loaded resonators. We consider the so-called round robin rotation method versus the single measurement technique and we evaluate the respective uncertainties. This comparison is done in the case of samples with surface resistance near that of the resonant cavity walls. We establish the limits of application of the round robin technique versus those of the single measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.