Abstract

We propose the collinear balanced detection (CBD) technique for noise suppression in fiber laser (FL)-based stimulated Raman scattering (SRS) microscopy. This technique reduces the effect of laser intensity noise at a specific frequency by means of pulse splitting and recombination with a time delay difference. We experimentally confirm that CBD can suppress the intensity noise of second harmonic (SH) of Er-FL pulses by 13 dB.The measured noise level including the thermal noise is higher by only ~1.4 dB than the shot noise limit. To demonstrate SRS imaging, we use 4-ps SH pulses and 3-ps Yb-FL pulses, which are synchronized subharmonically with a jitter of 227 fs. The effectiveness of the CBD technique is confirmed through SRS imaging of a cultured HeLa cell.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.