Abstract

A micro-machined thermal flow sensor to measure a minute amount of gas flow rate is developed, and tested to quantify the flow measuring characteristics over the variation of flow rate. A bifilar-Evanohm-R alloy heater and chromel-constantan thermopiles were formed on a sandwich membrane, Si 3N 4/SiO 2/Si 3N 4, which consist a core part of the sensor. In performance test of the sensor, range of gas flow rate was regulated from 0.2 to 1.0 slm and the heater power was variated from 0.72 to 5.63 mW. In the present experiment, the Reynolds number based on the diameter of a flow passage was ranged from 81 to 408. It is confirmed from the experiment that the sensitivity ( S = ∂Δ V/∂ Q; Δ V, voltage difference between thermopiles; Q, flow rate) of the sensor is enhanced with increasing the heater power and decreasing the measuring flow rate range. Also, the sensor response time required to a stable state is reduced with increasing the flow rate range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.