Abstract

LiF TLD-100 ribbons are frequently used for a wide range of radiobiological dosimetry applications using various sources in the dose range 0.1-10 Gy. It is suggested that an accuracy of 5-6% and a precision of 2-3% is required for the determination of absorbed dose in biological applications. To minimise the random errors, the dosemeters are calibrated individually. The authors examine the systematic errors that may be introduced by this procedure. They have observed, for instance, that the calibration procedure itself enhances the sensitivity of the dosemeters by 1-2% and that this enhancement is dependent upon their thermal history. They describe the procedures that have been established at the Gray Laboratory for routine precision thermoluminescence dosimetry. Particular attention is given to the sensitivity correction factor that is necessary to allow for enhancement effects due to supralinearity and the dosemeter thermal history.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.