Abstract

In conventional electrical capacitance tomography (ECT) systems the single-electrode excitation scheme is used in which an excitation signal is applied to each of the sensor electrodes in turn. This conventional ECT sensor is more sensitive in the wall area than the central area and consequently the signal-to-noise ratio (SNR) in the central area is poor. To improve this situation, a scheme called parallel field excitation has recently been proposed. This paper analyses the sensitivity distributions of the parallel field sensors. Analytical, simulation and experimental results show that the sensitivity maps of the parallel field sensors are simply the linear superimpositions of those of the conventional sensor, and give inferior image reconstructions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.