Abstract
Lumbar total disc arthroplasty (TDA) is an attractive option for the treatment of spinal disorders. However, post-operative complications may cause revision surgery. In this study, an in-depth biomechanical analysis was performed to test how sensitive post-operative spinal responses in sagittal bending were to design parameters of a widely used ball-in-socket TDA design. Our simulation results revealed that lowering the intervertebral distraction, posteriorly positioning the implant, or decreasing the radius of curvature of the metal-on-polyethylene bearing surface could result in spinal responses which better matched normal spinal kinematics, but deviated further from the normal spinal tissue load-sharing pattern.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Computer Methods in Biomechanics and Biomedical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.