Abstract
A simple, sensitive, accurate and affordable spectrofluorimetric method was developed and validated for the determination of venlafaxine, both in marketed preparations as well as in spiked rat plasma. Venlafaxine depicted strong native fluorescence property in freshly prepared 0.05 M sulphuric acid. The excitation and emission wavelengths were found to be 237.0 nm and 301.0 respectively. Effect of variations in pH, temperature, concentration, change in molarities of different solvents, and effect of excipients were studied. The calibration graph in case of dosage forms and in spiked plasma was found to be rectilinear in the concentrations of 15-600 ng/ml and 20-650 ng/ml respectively. The intra- day and inter-day accuracy measurements of VEN in formulations ranged from 0.29 to 0.44% and 0.27 to 0.49%, respectively. The intra-day and inter-day accuracy in measurement of VEN in plasma ranged from 0.062 to 2.26% and 0.52 to 2.32%, respectively. The limit of detection (LOD) was found to be 6.0 ng/mL and 4.0 ng/mL in plasma and formulations respectively. The mean recovery of VEN from plasma was 97.46.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.