Abstract

A simple, rapid, and ultra sensitive dispersive solid phase extraction based on nano graphene oxide was developed for simultaneous measurement of trace amounts of metformin (MET) and linagliptin (LIN) in plasma samples by HPLC-UV-Vis. Affecting factors on the extraction of these drugs, including adsorbent weight, extraction time, organic solvent type, desorption situations, and composition of solvent were examined and optimized. In optimum conditions, the LOD (limit of detection) and LOQ (limit of quantification) of the suggested technique were 2.0 ngmL-1 and 6.1 (ngmL-1) for LIN and 3.0 ngmL-1 and 9.2 ngmL-1 for MET, respectively. Suitable linear behavior in the considered ranges of concentration (10-2000 ngmL-1) and good correlation coefficient of 0.9901 and 0.9903 (r2) for LIN and MET were obtained, respectively. The RSD (relative standard deviations) according to three replicate measurements at 2, 20, 200 ngmL-1 levels of these drugs was less than 8.0%. In the last step, applicability of the suggested technique was examined by analyzing the drugs in plasma samples and reasonable results were achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.