Attenuated total reflection spectrometry with a slab optical waveguide (SOWG) was explored for the simple, rapid and sensitive measurement of total anionic surfactants by the methylene blue active substance (MBAS) method. A fused-silica sheet used as a guiding layer was modified with trimethylsilane (TMS) to extract and concentrate the MBASs on the SOWG surface. Based on preliminary studies of the adsorption behavior and visible ATR spectrum of MB on the modified silica surface, a detection wavelength of 600 nm was chosen for the sensitive measurement of anionic surfactants. When the concentration of MB was set at 10 microM in the final measurement solution, the calibration curve for a typical anionic surfactant (sodium dodecylbenzenesulfonate) was linear up to 0.6 microM and the detection limit was 0.07 microM. The proposed method was applied to the determination of total anionic surfactants in river water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.