Abstract
The development of an automated kinetic-catalytic spectrophotometric method for cobalt determination is presented. The method is based on the catalytic effect of Co in the oxidation of hydroxybenzoic acid by H2O2 in basic media. The method has been automated using a multisyringe flow injection system coupled to a monolithic flow conduit called chip (chip-MSFIA). All reagents and sample are simultaneously propelled into the chip to achieve an efficient mixing. The reaction product is monitored at 482nm. The reaction takes place very fast at room temperature, thus the fixed-time method is applied to quantify Co concentration in samples. Variables such as, reagents concentration, pH, flow rate and reaction time have been optimized to improve the selectivity and sensitivity of the proposed system. Under optimal conditions, Co may be determined in the range 0.02–10.00μgL−1 achieving a limit of detection of 0.02μgL−1 and an injection throughput of 68h−1. Relative standard deviations are below 3%. The method has been successfully applied to water samples and a pharmaceutical formulation. The accuracy of the method has been validated by add-recovery tests and satisfactory recoveries from 91% to 97% were obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.