Abstract
A rapid and sensitive high-performance liquid chromatography (HPLC) method was developed to detect perchlorate in tissues of male and female rats, both pregnant and lactating (including milk) after administration of perchlorate. Supernatants of ethanol precipitated rat fluids and tissues were evaporated to dryness under nitrogen and reconstituted in deionized water. Reconstituted samples were injected into HPLC system coupled with conductivity detection. Isocratic separation of perchlorate was achieved using an anion-exchange column with sodium hydroxide as mobile phase and a conductivity detector. In this method, perchlorate showed a linear response range from 5 to 100 ng/ml. The lower detection limits for perchlorate in fluids and tissues of rats were 3–6 ng/ml and 0.007–0.7 mg/kg, respectively. The described method has the unique advantage over the existing methods of determining low traces of perchlorate in different biological matrices without complex sample preparation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.