Abstract

A simple and sensitive method is presented for determination of styrene, toluene, ethylbenzene, isopropylbenzene andn-propylbenzene in human body fluids by capillary gas chromatography (GC) with cryogenic oven trapping. After heating a blood or urine sample containing each compound andp-diethylbenzene (internal standard, IS) in a 7.0-mL vial at 60°C for 20 min, 5 mL of headspace vapor was drawn into a glass syringe and injected into a GC. All vapor was introduced into an Rtx-Volatile middle bore capillary column in splitless mode at oven temperature of 20°C to trap entire analytes, and the oven temperature then programmed to 280°C for GC measurements by flame ionization detection. The present conditions gave sharp peaks of each compound and IS, and low background noises for whole blood or urine samples.