Abstract

A reliable and sensitive method for simultaneous determination of bisphenol A (BPA) and bisphenol F (BPF) in canned food by gas chromatography-mass spectrometry (GC/MS) is described after extraction and pre-concentration by a new solid-phase microextraction (SPME) adsorbent. The potential of single-walled carbon nanotubes (SWCNTs) as SPME adsorbent for the pre-concentration of environmental contaminants has been investigated in recent years. This work was carried out to investigate the feasibility of SWCNTs as a headspace SPME adsorbent for the determination of bisphenol derivatives in canned food. Potential factors affecting the extraction efficiency, including extraction time, extraction temperature, desorption time, desorption temperature, and salinity were optimized. Calibration curves were linear (r 2≥ 0.994) over the concentration range from 0.30 to 60 µg kg−1. For both target analytes, the limit of detection (LOD) at signal-to-noise (S/N) ratio of 3 was 0.10 µg kg−1. In addition, a comparative study between the SWCNT and a commercial polydimethylsiloxane (PDMS) SPME fibre for the determination of bisphenol derivatives in canned food was conducted. SWCNT fibre showed higher extraction capacity, better thermal stability (over 350°C) and longer life span (over 150 times) than the commercial PDMS fibre. The method was successfully applied to determine BPA in canned food samples which were purchased from local markets. BPA was found in some of the samples within the concentration range from 0.5 to 5.2 µg kg−1.