Abstract
This paper introduces a simple method for selective and sensitive detection of black powder by adding chlorinated hydrocarbons in the drift gas instead of changing the structure of conventional ion mobility spectrometer (IMS). The function of chloride modifiers was to substitute Cl(-)(H₂O)n for [O₂⁻ (H₂O)(n)] in the drift region so as to avoid the overlap between O₂⁻ (H₂O)(n) and sulfur ion peaks. Among CH₂Cl₂, CHCl₃ and CCl₄, CCl₄ was chosen as the modifier due to the best peak-to-peak resolution and stability towards the fluctuation of modifier concentration. With 1.4 ppm CCl₄ as the modifier, the minimum detectable quantity of 0.1 ng for sulfur was achieved. Moreover, this method showed the ability for detection of common explosives at sub-nanogram level, such as black powder (BP), ammonium nitrate fuel oil (ANFO), 2,4,6-trinitrotoluene (TNT), and pentaerythritol tetranitrate (PETN). In summary, this method requiring no configuration modification has high sensitivity and selectivity, and consumes trace amount of modifier. And these characteristics make it easy to be adopted in current deployed IMS to detect black powder explosives.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.