Abstract

Abstract. Proton-transfer-reaction mass spectrometry (PTR-MS) is a technique that is widely used to detect volatile organic compounds (VOCs) with proton affinities higher than water. However, n-alkanes generally have a lower proton affinity than water and therefore proton transfer (PT) by reaction with H3O+ is not an effective mechanism for their detection. In this study, we developed a method using a conventional PTR-MS to detect n-alkanes by optimizing ion source and drift tube conditions to vary the relative amounts of different primary ions (H3O+, O2+, NO+) in the reaction chamber (drift tube). There are very few studies on O2+ detection of alkanes and the mixed mode has never been proposed before. We determined the optimum conditions and the resulting reaction mechanisms, allowing detection of n-alkanes from n-pentane to n-tridecane. These compounds are mostly emitted by evaporative/combustion process from fossil fuel use. The charge transfer (CT) mechanism observed with O2+ was the main reaction channel for n-heptane and longer n-alkanes, while for n-pentane and n-hexane the main reaction channel was hydride abstraction (HA). Maximum sensitivities were obtained at low E ∕ N ratios (83 Td), low water flow (2 sccm) and high O2+ ∕ NO+ ratios (Uso = 180 V). Isotopic 13C contribution was taken into account by subtracting fractions of the preceding 12C ion signal based on the number of carbon atoms and the natural abundance of 13C (i.e., 5.6 % for n-pentane and 14.5 % for n-tridecane). After accounting for isotopic distributions, we found that PT cannot be observed for n-alkanes smaller than n-decane. Instead, protonated water clusters of n-alkanes (M ⋅ H3O+) species were observed with higher abundance using lower O2+ and higher water cluster fractions. M ⋅ H3O+ species are probably the source for the M + H+ species observed from n-decane to n-tridecane. Normalized sensitivities to O2+ or to the sum of O2++ NO+ were determined to be a good metric with which to compare sensitivities for n-alkane detection between experiments. Double hydride abstraction was observed from the reaction with O2+. Sensitivity to CT increased with carbon chain length from n-pentane to n-dodecane, sensitivity to HA increased from n-heptane to n-dodecane and sensitivity to PT increased from n-decane to n-tridecane. Sensitivity to CT exponentially decreased with molecular ionization energy, which is inversely related to the carbon chain length. We introduce a calibrated fragmentation algorithm as a method to determine the concentrations of n-alkanes and demonstrate its effectiveness using a custom n-alkane mixture and a much more complex oil example representing perhaps the most difficult mixture available for application of the method. We define optimum conditions for using the mixed ionization mode to measure n-alkanes in conventional PTR-MS instruments regardless of whether they are equipped with switchable reagent ion (SRI) capabilities.

Highlights

  • Proton-transfer-reaction mass spectrometry (PTR-MS) has been extensively used to measure volatile organic compounds (VOCs) (e.g., de Gouw and Warneke, 2007)

  • Under normal PTR-MS conditions optimized for H3O+, there are still O+2 and NO+ ions produced as byproducts in the ion source discharge

  • We explored and optimized mixed ionization mode conditions in the PTR-MS to enhance the detection of n-alkanes

Read more

Summary

Introduction

Proton-transfer-reaction mass spectrometry (PTR-MS) has been extensively used to measure volatile organic compounds (VOCs) (e.g., de Gouw and Warneke, 2007). O. Amador-Muñoz et al.: Sensitive detection of n-alkanes other species that are extremely important for air quality, atmospheric chemistry and source emission characterization. Amador-Muñoz et al.: Sensitive detection of n-alkanes other species that are extremely important for air quality, atmospheric chemistry and source emission characterization This technique offers close to real-time data acquisition and low detection limits (low pptv) (Karl et al, 2012). Chemical ionization (CI) from different reagent ions is possible in PTR-MS instruments equipped with switchable reagent ions (SRI) and with selected-ion flow tube mass spectrometers (SIFT-MS). The proton affinities for commonly encountered inorganic compounds in air, such as nitrogen and oxygen, and for some haloalkanes and alkanes, are higher and they are not detectable by PT (Reaction R1)

Methods
Results
Conclusion