Abstract
© 2017 The Authors. The present work describes the sensitive and low-potential detection of hydroquinone (HQ) using a glassy carbon electrode (GCE) modified with nanodiamond (ND). The presence of ND was confirmed by high-resolution scanning electron microscopy, FTIR and Raman spectroscopy. The cyclic voltammetry results reveal that ND modified GCE has high electrocatalytic activity towards oxidation of HQ than unmodified GCE. In addition, the ND modified GCE shows a significantly lower oxidation potential towards CC than unmodified GCE, which is due to the presence of -OH functional groups in ND. The CV studies confirm that the redox electrochemical behavior of CC is a diffusion controlled electrochemical process on ND modified electrode. The ND modified electrode was able to detect the HQ across a linear response range from 1.0 to 78.0 μM. The limit of detection of the sensor was 0.19 μM. The sensor shows an appropriate selectivity in the presence of 20-fold concentrations of benzenediols and neurotransmitters. The practicality of the sensor is appropriate and shows excellent recovery of HQ in lab water samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.