Abstract

International guidelines suggest a screening panel for monoclonal gammopathies that contains serum protein electrophoresis (SPE), free light chain (FLC) measurements and immunofixation. This combination provides the possibility of a timely accurate diagnosis. To evaluate the sensibility of a simple screening panel (SPE + FLC). We analyzed 191 consecutive serum samples of patients with a suspected monoclonal gammopathy (MG). Seventy five patients were diagnosed with MG. The sensitivity and specificity of the combination of SPE + FLC for the diagnosis of monoclonal gammopathy were 95% (95% confidence intervals 89-99) and 99% (95% confidence intervals 96-100), respectively. We were able to validate the international recommendations on the diagnostic accuracy of this simple combination of two tests in serum for monoclonal gammopathy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call