Abstract
The determination of beryllium using electrothermal atomic absorption spectrometry with deuterium background correction in the presence of various isomorphous metals and Mg(NO 3) 2 was studied. While, Eu, Ir and Sm had no effect on the transient signals, the addition of Lu and Mg(NO 3) 2 improved the sensitivity of the beryllium signal with respect to that obtained in the absence of modifier. Although, Mg(NO 3) 2 has improved the signal with respect to its sensitivity, it also increased the tail and the background (BG) signals, specially when urine samples are under study. Whereas, when Lu was used the analytical signal is virtually free of BG interference indicating that the urine matrix interference was almost eliminated. Besides, the addition of 6 μg of Lu ensured that the signals were effectively constant for five firings following the furnace program, which included: three drying, and the pyrolysis, atomization, cleaning and cooling steps. The effect of some components, likely to interfere in the accurate determination of beryllium (such as: Al, Ca, Cl, Co, Cr, Fe, Mg and Mn) were investigated. At the physiological levels, most of these elements had no effect, except in the case of chloride when Mg(NO 3) 2 was used as modifier. In this case, the tolerance limit was of 3000 mg Cl − l −1. The characteristic masses were 1.19, 0.45 and 0.48 pg, when integrated absorbance was measured for beryllium without the addition of any modifier and in the presence of Lu and Mg(NO 3) 2, respectively. The limits of detection (3σ) were 85, 19 and 58 fg, respectively. The accuracy and precision with the use of Lu and Mg(NO 3) 2 was tested for the direct determination of beryllium in urine samples. Quantification was performed with aqueous standards. The results obtained for the determination of beryllium in reference materials (Trace Elements in Urine), together with good recovery of spiked analyte, using either Lu or Mg(NO 3) 2 modification demonstrate the applicability of the procedure to the analysis of real samples. However, Lu provided the most accurate results. Also, the addition of Lu enhanced the precision of the measurements to levels of 1.8% relative standard deviation instead of 5.6 and 3.3% for the case of beryllium alone and with the addition of Mg(NO 3) 2.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.