Abstract
A technique is described for performing solid-phase extractions on a semi-micro scale. Thin membrane disks 4 mm in diameter containing lightly sulfonated polystyrene or Silicalite particles are placed in the hub of a syringe needle. Aqueous samples (1–6 ml) are passed through the membrane disks and the extracted compounds are subsequently eluted with 20–50 μl of an organic solvent. Unlike solid-phase micro extraction (SPME) which uses a coated fiber, the present method is essentially a total extraction technique. Recoveries >90% were generally obtained for a wide variety of test compounds. The same test compounds in human urine, albumin and human serum samples can be extracted without any pretreatment other than addition of a suitable surfactant. A “double-pass” technique was developed for convenient field sampling.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.