Abstract

A procedure is described for the determination of carbon by closed flask combustion. It involves electrical ignition of the sample (15–20 mg) wrapped in a high melting glass fiber, absorption of the CO2 in NaOH-BaCl2 solution, separation and dissolution of the BaCO3 formed in HCl and back titration with NaOH. This method has been tested with organic compounds containing nitrogen, sulfur and chlorine.

Full Text