Abstract

The possibility of measurement of electrons ionization loss in Si layer of smoothly tunable thickness is shown in the proof-of-principle experiment. The Si surface-barrier detector with the depleted layer thickness controlled by the value of high voltage power supply has been used. Ionization loss spectra for electrons emitted by radioactive source 207Bi are presented and discussed. Experimental results for the most probable ionization loss in the Landau spectral peak are compared with theoretical calculations. The possibility of research of evolution of electromagnetic field of ultra-relativistic particles traversing media interface with the use of detectors with smoothly tunable thickness is proposed.