Abstract

A new approach using a chromatography system equipped with isocratic pumps and an electrolytic eluent generator (EG) is introduced, replacing external pH gradient delivery using conventional gradient systems, in which bottled buffers with preadjusted pH are mixed using a gradient pump. The EG is capable of generating high purity base or acid required for online preparation of the buffer at the point of use, utilizing deionized water as the only carrier stream. Typically, the buffer was generated from online titration of a reagent composed of low molecular weight amines. The reagent was delivered isocratically into a static mixing tee, where it was titrated to the required pH with electrolytically generated base or acid. The required pH gradient was thus conveniently generated by electrically controlling the concentration of titrant. Also, since the pH was adjusted at the point of use, this approach offered enhanced throughput in terms of eluent preparation time and labor, and with a more reproducible pH profile. The performance of the system was demonstrated by running pH gradients ranging from pH 8.2 to 10.9 on a polymer monolith cation-exchange column for high throughput profiling of charge heterogeneity of intact, basic therapeutic monoclonal antibodies. A high degree of flexibility in modulating the key parameters of the pH gradient, including the buffer concentration, the pH gradient slope and the operating pH range was demonstrated. This enabled fine-tuning of the separation conditions for each individual antibody in order to enhance the chromatographic resolution.