The paper describes semi-automated procedure for determination of 89,90Sr in drinking water. It includes quantitative capture of 89,90Sr on SuperLig 620 column from large volume of water samples and their on-column detection via Cherenkov counting on liquid scintillation counter. In this case, SuperLig 620 column acts as a specific strontium separator and a sensor for Cherenkov counting. The method enables determination of 89Sr and 90Sr above 30 mBq L−1 with good accuracy within 64 h.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.