Abstract

This semi-automated colorimetric assay for the MB isoenzyme of creatine kinase (EC 2.7.3.2) is based on a monoclonal antibody ("Conan-MB") specific for this isoenzyme and is a modification of a previously published method (Vaidya et al., Clin Chem 1986;32:657-63). A 0.64-cm bead coated with 2 to 3 micrograms of antibody is incubated with 100 microL of serum and 10 microL of 0.2 mol/L beta-mercaptoethanol for 1 h at room temperature, to extract CK-MB. The beads are washed with de-ionized water and incubated with CK substrate for 45 min at 37 degrees C. A solution containing trans-1,2-diaminocyclohexane-N,N,N', N'-tetraacetic acid, p-iodonitrotetrazolium violet, and diaphorase is added and the resulting colored product is measured at 492 nm. The standard curve is linear to 200 U of CK-MB per liter, and analytical recovery is 97-113%. Total assay CV for low (9.7 U/L) and high (50.7 U/L) quality-control materials was 14.1% (n = 1878) and 11.6% (n = 1842), respectively. CK-MB activity correlated well (r = 0.978, n = 226) with CK-MB measured by a two-site mass immunoassay, and 99.4% of samples with CK-MB greater than or equal to 12 U/L (n = 347) were verified by electrophoresis on agarose.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.