Abstract

Numerous discarded ProFile GT, ProFile, and ProTaper nickel-titanium rotary instruments obtained from two graduate endodontic clinics were examined by scanning electron microscopy. These instruments had an unknown history of clinical use and had fractured or experienced considerable permanent torsional deformation without complete separation. The failure processes generally exhibited substantial ductile character, evidenced by a dimpled rupture fracture surface. Crack propagation at grain boundaries and cleavage surfaces indicative of transgranular fracture were observed for some specimens. It appeared that oxide particles from the manufacturing process served as nucleating sites for the microvoids, leading to dimpled rupture. A previously unreported fracture mode also was observed, in which crack propagation, approximately parallel to the local flute orientation, connected pitted regions on the surface. Combining present and previous scanning electron microscopy observations of clinically failed instruments, suggestions are offered for improving their fracture resistance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.