Abstract
Self-absorption was measured for the activity calculation of the 1460.8 and 2614.47keV lines of the 40K and the 208Tl, respectively, in powdered milk samples. Five Marinelli beakers were prepared with powdered milk in different degrees of compaction and the spectra were measured with an HPGe detector. The detection efficiency versus density was obtained and the self-absorption correction factors versus density were calculated for powdered milk. The results obtained show that this factor must be considered in the calculations of activities.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.