Abstract
The intensity of the hyperfine doublet of the Cu(I) 324.8-nm transition was observed as a function of current, fill gas, and fill gas pressure for a new design of hollow cathode lamp incorporating a positive column and an enlarged volume cathode cavity. The intensity ratio of the two components, which reflects the degree of self-absorption, and the flame absorption sensitivity were compared with those of a conventional hollow cathode lamp. Higher intensity, better component ratio, and increased absorption sensitivity were found in the new lamp.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.