Abstract

The intensity of the hyperfine doublet of the Cu(I) 324.8-nm transition was observed as a function of current, fill gas, and fill gas pressure for a new design of hollow cathode lamp incorporating a positive column and an enlarged volume cathode cavity. The intensity ratio of the two components, which reflects the degree of self-absorption, and the flame absorption sensitivity were compared with those of a conventional hollow cathode lamp. Higher intensity, better component ratio, and increased absorption sensitivity were found in the new lamp.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.