Electrical Impedance Tomography (EIT) is a medical imaging technique which has the potential to reduce time to treatment in acute stroke by rapidly differentiating between ischaemic and haemorrhagic stroke. The potential of these methods has been demonstrated in simulation and phantoms, it has not yet successfully translated to clinical studies, due to high sensitivity to errors in scalp electrode mislocation and poor electrode-skin contact. To overcome these limitations, a novel electrode helmet was designed, bearing 32 independently controlled self-abrading electrodes. The contact impedance was reduced through rotation on an abrasive electrode on the scalp using a combined impedance, rotation and position feedback loop. Potentiometers within each unit measure the electrode tip displacement within 0.1 mm from the rigid helmet body. Characterisation experiments on a large-scale test rig demonstrated that approximately 20 kPa applied pressure and 5 rotations was necessary to achieve the target 5 kΩ contact impedance at 20 Hz. This performance was then replicated in a simplified self-contained unit where spring loaded electrodes are rotated by servo motors. Finally, a 32-channel helmet and controller which sequentially minimised contact impedance and simultaneously located each electrode was built which reduced the electrode application and localisation time to less than five minutes. The results demonstrated the potential of this approach to rapidly apply electrodes in an acute setting, removing a significant barrier for imaging acute stroke with EIT.

Highlights

  • Thrombolytic agents can only be used to treat ischaemic strokes as they are potentially fatal for patients with haemorrhages, so it is vital that cause of stroke is identified prior to treatment

  • The results suggest that without paste the impedance has decreased greater than the equivalent of 6 tape strippings or the removal of 48 μm of the Stratum Corneum, whereas abrasion with paste improved upon manual abrasion with impedances lower than observed after 9 strippings or 72 μm removed

  • The results demonstrated that so long as the torque applied was sufficient to overcome friction, the effective reduction of contact impedance was dependent predominantly on the total angle rotated for a wide range of applied pressures

Read more

Summary

Introduction

Stroke is a leading cause of morbidity and disability in industrialised nations [1]. It is the third leading cause of lost life years [2] and estimated that total aggregate annual costs of stroke to the UK is. Thrombolytic agents can only be used to treat ischaemic strokes as they are potentially fatal for patients with haemorrhages, so it is vital that cause of stroke is identified prior to treatment. This cannot be achieved without neuroimaging, and in the UK, the National Institute for Health and Clinical

Methods
Findings
Discussion
Conclusion