Abstract
The chromatographic behaviour of a commercially available ion-exchange stationary phase (the Dionex IonPac CS12A column) is described for a wide range of transition and heavy metal ions with nitric acid eluents containing chloride and nitrate potassium salts. The separation selectivity was found to arise from simultaneous ion-exchange interactions and chelation with the attached carboxylic and phosphonic acid groups. These interactions were investigated by altering the ionic strength and pH of the eluent and also the column temperature. Strong affinity of the stationary phase towards heavy metal ions, in particular bismuth and the uranyl ion was observed at low pH under chelating ion-exchange conditions, with high efficiency separations of other ions including cadmium and lead being possible with short analysis times (∼5–15 min). Examples are given of separations obtained using 4-(2-pyridylazo)resorcinol or Arsenazo III as the post-column chromogenic reagents, demonstrating the potential versatility and utility of this stationary phase for heavy metal ion analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.