Abstract
A multi-contact cylindrical nerve cuff electrode was evaluated for its ability to record neural signals selectively in an in vitro preparation, Three branches of a Beagle hypoglossal nerve are stimulated sequentially while compound action potentials (CAP) are recorded from its trunk with the multi-contact cuff electrode. A selectivity index (SI) is defined and applied to the CAP recorded from the 4 sets of tripolar contacts (12 contacts in total) that are equally spaced around the cuff. The results show that the cuff can record selectively from different fascicles, but these effects are small. Connecting the contacts of the opposite set together while recording from a tripole improved the selectivity. The signal amplitudes from various contacts are consistent with the location of the fascicles relative to the contacts.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.