Abstract

A selective and sensitive liquid chromatographic method was developed for the determination of fluoxetine (FLU) in plasma. FLU was isolated from plasma by liquid-liquid extraction. The chromatographic separation was performed on an analytical 250 x 3.9 mm id Novapak C18 column (4 microm particle size) with an isocratic mobile phase consisting of phosphate buffer-acetonitrile-methanol -triethylamine (58 + 30 + 10 + 2, v/v) adjusted to pH 7. Using UV detection at 226 nm, the detection limit for FLU in plasma was 3 ng/mL. No interferences were found with tricyclic antidepressant drugs, which allows this method to be used in clinical studies. The calibration curve was linear over the concentration range of 10-200 ng/mL. The average recovery was about 80% for plasma. The inter- and intraday assay coefficients of variation were <8%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.