Abstract

AbstractThe frequent detection of physiological glucose levels from human blood or sweat requires the development of low‐cost electrodes with high sensitivity and selectivity. In this work, we prepared a series of Prussian blue (PB) modified carbon cloth (CC) electrodes with different cyanoferrate groups. We achieved a sensitivity as high as 145.43 μA mm−1cm−2 in a 0.1–6.5 mm concentration range with a response time below 2 s under physiological pH. The electrodes exhibited a superior selectivity of glucose in the presence of interfering agents, including sucrose, lactose, NaCl, ascorbic acid, and uric acid. The electrodes also showed outstanding long‐term stability over 15 days. Furthermore, we performed comprehensive electrochemical and characterization studies to elucidate the role of the cyanoferrate group on the morphologic and electronic properties of non‐enzymatic glucose sensors.