Abstract

Capillary zone electrophoresis was employed for the analysis of cytochrome c using an end-column amperometric detector with a L-cysteine-modified gold microdisk array electrode. The optimum conditions of separation and detection are 0.04 mol/L KH2PO4 - 0.06 mol/L Na2HPO4 for the buffer solution, 8 kV for the separation voltage, 4 kV and 10 s for the injection voltage and the injection time, and −0.12 V vs. SCE for the detection potential. When the fused-silica capillary modified by ethylene glycol was used, the adsorption of cytochrome c on the surface of the capillary wall could be overcome. The limit of detection is 3 × 10−5 mol/L or 62 fmol (S/N = 2). The method was applied to determining cytochrome c in pig heart.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.