Abstract

An ultrasensitive voltammetric sensor has been developed for the simultaneous quantification of amlodipine (AML) and lisinopril (LSP) on PANI-CeO2/gold sensor. The experimental parameters, such as concentration, pH, amplitude, frequency, and deposition potential were optimized.AML and LSP have shown two well defined reduction peaks and the cathodic peak potential were found to be −0.4 V for LSP and −0.8 V for AML, respectively by square-wave voltammetry (SWV). The limits of detection (LOD) were found to be 0.036 nmolL−1 for LSP and nmolL−1 for AML respectively. The proposed method was applied to simultaneously estimate the amount of drug in pharmaceutical formulation with almost 99.0% recovery and in spiked human serum with good recovery of almost 98.0%. Developed method showed better LOD as compared to other techniques.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.